| The Cleo is a new and intuitive approach to patient vital signs measurement. The Cleo can be configured to measure any combination of: non-invasive blood pressure, SpO2, rapid temperature, and capnography (EtCO2). |
| |
| Weighing in at less than 3 LBS the portable Cleo is well suited for any patient care area by offering a multitude of vital sign combinations. The Cleo can be used as a basic pulse oximeter or configured to a NIPB/SpO2/Temp spot check monitor. Cleo can also be configured to be a stand-alone capnography or combination capnography/SpO2/NIPB monitor. The Cleo is well suited for both bed side and mobile spot check use. |
| |
| Specifications: |
| Neonatal, pediatric and adult patients |
| |
| PHYSICAL DIMENSIONS & WEIGHT : |
| Base Unit: 8 x 4.5 x 4 (HxWxD inches) |
| Weight: 2.5 LBS |
| |
| PERFORMANCE SPECIFICATIONS: |
| Display: 5.0 inch (Diagonal) color TFT |
| Resolution: 800 × 3(RGB) × 480 |
| Trace: 2 waveforms |
| Waveforms: PLETH, ETCO2 |
| Indicator: Alarm Indicator |
| Power indicator |
| Pulse beep and alarm sound |
| Trend time: From 1 to 72 hours |
| NIBP |
| Measuring Technology: Automatic oscillating measurement |
| Cuff Inflating: <30s (0 ~ 300 mmH, standard adult cuff) |
| Measuring Period: AVE<40s |
| Mode: Manual, Auto, STAT |
| |
| Measuring Interval: |
| in AUTO Mode: 2 min ~ 4 hrs |
| Pulse Rate Range: 30 bpm ~ 250 bpm |
| Measuring Range: Adult/Pediatric Mode |
| SYS: 40 ~ 250 (mmHg) |
| DIA: 15 ~ 200 (mmHg) |
| |
| Neonatal Mode: |
| SYS: 40 ~ 135 (mmHg) |
| DIA: 15 ~ 100 (mmHg) |
| Resolution: 1mmHg |
| Pressure Accuracy: Maximum Mean error: ±5mmHg |
| Maximum Standard |
| deviation: 8mmHg |
| Overpressure Protection: Adult Mode: 280(mmHg) |
| Neonatal Mode: 150 (mmHg) |
| Alarm Limit: SYS: 50 ~ 240 mmHg |
| DIA: 15 ~ 180 mmHg |
| Standards: Meets performance standards of |
| ANSI/AAMI SP10:2002 |
| SPO2 |
| ASpO2: Anti-motion SpO2 |
| SpO2% Range: 0 ~ 100% |
| SpO2 Accuracy: ±2% (70 ~ 100%,non-motion) |
| ±3% (70 ~ 100%, motion) |
| Pulse Rate Range: 30-250 bpm |
| Pulse Rate Accuracy: ±2 bpm(non-motion), |
| ±3 bpm (motion) |
| Alarm Upper-lower Limit: Upper limit 70 ~ 100%, |
| Lower limit 70 ~ 100% |
| SpO2 Probe: Red light LED wavelength |
| 660nm±5nm |
| Infrared light LED wavelength: |
| 940nm±10nm |
| Standards: Meets performance standards |
| of EN ISO 9919:2005 |
| RAPID TEMPERATURE (OPTION) |
| |
| Temperature: |
| Seconds, 2 “beeps” at 30 seconds, |
| 1 “beep” at 45 seconds, and 2 |
| “beeps”at 60 seconds |
| Patient Accuracy: A Standard Prediction Mode reading |
| and a Direct Mode reading will differ |
| by less than ±0.2°C (±0.4°F) on 98% |
| of tested patients |
| Batteries: Four “AA” Required. |
| Standard IEC package size. |
| Alkaline --1.5 Volt |
| Approx. 6000 temperature readings |
| EtCO2 (OPTION) |
| Mode of Sampling: Sidestream or Mainstream |
| Principle of Operation: Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts. |
| CO2 measurement Range: 0 to 150 mmHg breath |
| |
| Response Time:: |
| <3 seconds - includes transport time and rise time |
| Inspired CO2 |
| measurement Range: 3 ~ 50 mmHg |
| Standards: Meets performance standards of ISO/ |
 |